If your engravings look soft or inconsistent, your autofocus may not match your manual focus gauge. This guide shows you how to find true focus on your Epilog Mini 18/24 or Helix, set the manual gauge to it, and then adjust autofocus so both agree.
Applies to
- Epilog Mini 18
- Epilog Mini 24
- Epilog Helix
What you need
- 4″ × 4″ piece of anodized aluminum
- #1 Phillips screwdriver
- Manual focus gauge (the V-shaped tool)
- CorelDRAW or other graphics software, with an Epilog print driver
Part 1: Set up the test job
Part 2: Find true focus and set the manual gauge
5. Stop the job
Press Stop.
Don’t press Reset. Reset moves the table and you’ll lose the focus height you just found.
Part 3: Match autofocus to the manual gauge
8. Adjust the value and save
Use the arrows to change the value, then press Go to save it.
- Range: −200 to +300
- More negative: the table sits higher
- More positive: the table sits lower
- 40 units: about 1/16″ of table movement
9. Run the job again
Run the job and press Stop partway through, so the table stays at the autofocus height.
10. Check against the gauge
Hang the manual gauge in place and look at the tip:
- Just touching the aluminum: you’re done. Autofocus and manual focus now match.
- Gap under the tip: the table is too low. Make the value more negative.
- Pressing into the aluminum: the table is too high. Make the value more positive.
Repeat steps 7–10 until it’s right. Leave the gauge screws alone during this part; you’re matching autofocus to the gauge, not the other way around.