This guide covers replacing the top door assembly on the Fusion and Fusion M2. The gas shocks and hinges are disconnected, the old door is removed and the new door is installed.
Parts Required
- Top Door Assembly
Tools / Materials Required
- 5/32" Allen Wrench
- 3/8" Stubby Nut Driver or 3/8" Short Socket Wrench
- Small Flat Head Screwdriver
- Permanent Marker (hinge plate machines)
- A second person
Safety First: Ensure the engraver is completely powered off and the power cord is disconnected before removing any panels or touching internal components.
Two People Required: The door is heavy. Do not remove or install it alone.
Before You Start
- Do not let the door fall while removing the old door or installing the new one.
- Mount the new door so that it closes fully and trips the interlock (safety) switches.
- The door is held at the hinges in one of two ways: a spring loaded lever on older Fusion machines, or a bolted hinge plate on newer Fusion and Fusion M2 machines.
Door Removal
Power Off Engraver
Turn off the engraver and unplug the power cord.
Remove the Upper Side Panels
Remove the upper left and upper right side panels. Each panel has two 5/32" Allen locks that only need a quarter turn (90°) counterclockwise to free the panel. Set the panels aside.
Note: Each side panel has two metal tabs at the bottom that lock it in place. Watch for these tabs when removing and replacing the panels.
Identify the Hinge Type
Find out which hinge type the engraver has, but do not remove anything yet.
- Spring loaded lever (older Fusion): pulling the lever out releases the door.
- Hinge plate (newer Fusion and Fusion M2): a metal locking plate held on with four 3/8" lock nuts.
Open the Door Fully
Open the door all the way. The gas shocks help support the weight of the door during the next steps.
Remove the Gas Shock Clips
Each end of each gas shock has a small metal C-shaped clip that holds it on its ball joint. While a second person holds the door up, pull the C clips off the door end of both shocks. A small flat head screwdriver can help.
Important: Once the shocks are off, the person holding the door is supporting almost its full weight, and these doors are heavy.
Remove the Shocks and Close the Door
Pull the gas shocks off their ball joints on the door. Then, preferably with two people, slowly close the door all the way. Let the shocks rest on the table or hang down.
Note: On a Fusion M2, open the front door before closing the top door fully.
Release the Door Hinges
Release the door at both hinges:
- Spring loaded lever (older Fusion): with one person on each side, pull the levers out to disconnect the door, slide the door forward a few inches and let go of the levers.
- Hinge plate (newer Fusion and Fusion M2): trace around both hinge plates with a permanent marker, remove the four 3/8" lock nuts on each plate and pull both plates out.
Remove the Old Door
With at least two people, lift the old door out of the engraver.
Important: The door is heavy, and very heavy on dual source machines.
Door Installation
Place the New Door
Set the new door in place:
- Spring loaded lever (older Fusion): place the door just in front of the spring loaded hinge pins.
- Hinge plate (newer Fusion and Fusion M2): place the door all the way into position. On a Fusion M2, the front door should still be open.
Reinstall the Hinges
Reconnect the door at both hinges:
- Spring loaded lever (older Fusion): with one person on each side, pull the levers out, slide the door all the way back and slowly let go of the levers. You may need to lift the back of the door slightly, slide it forward slightly, or both, to get the hinge pins into the new door.
- Hinge plate (newer Fusion and Fusion M2): line the door up as closely as possible with the hinge plate holes. Push the hinge plates back in, making sure they go into the new door, and install the four lock nuts on each plate. Leave them a little loose for now.
Reinstall the Gas Shocks
With at least one person holding the door up, press each shock onto its ball joint on the door, then slide the metal C clip back over the neck of the ball joint socket to lock it. Repeat for the other shock.
Set the Hinge Plates
Hinge plate machines only: the hinge plates are slotted so they can be set in the best position, which is why you traced around them. Lift the door, move each plate back to its traced outline and fully tighten the lock nuts on both plates.
Check the Door
Close the door and make sure it closes fully and trips the interlock (safety) switches.
Reinstall the Side Panels
Reinstall the upper left and upper right side panels, seating the tabs at the bottom first, and turn each Allen lock a quarter turn clockwise.