Focus and Camera Calibrations – Fusion Galvo

How do I calibrate the focus and camera system in the Fusion Galvo?

Introduction

Focus and camera calibrations keep the Fusion Galvo G100 at the correct focal height and keep the cameras accurate. Always complete the focus and camera calibrations together; skipping the camera calibration can leave the camera stream inaccurate and throw off artwork placement.

Parts Required

  • GC0208 – Target, Camera Calibration, G100

Tools / Materials Required

  • 8″ x 8″ Piece of Anodized Aluminum
  • 1/8″ Allen Wrench

When to Calibrate

Complete this procedure if you notice any of the following:

  • Laser power appears weak at the currently calibrated focal height
  • Galvo platform movement does not match focus input, or the platform crashes
  • Artwork placement with the IRIS camera system is inaccurate

It is also required after replacing any of these components:

  • Z-Axis PCB
  • Fiber laser
  • Control module
  • Galvo head assembly
  • Overhead camera assembly
  • Galvo lens

Note: The G100 uses two lenses: the F163 (4″ x 4″ field) and the F254 (6″ x 6″ field). Complete these calibrations with each lens installed. This guide shows the F254 lens; the F163 procedure is identical. Complete all six calibrations below in order, in one session.


Galvo Platform Calibration

The galvo platform moves from the top to the bottom of the z-axis, locating the magnetic sensors along the z-axis PCB. This step must be completed for an accurate focus calibration.

Step 1

Open Settings Menu

Press the gear icon to open the Settings menu.

Step 2

Open Focus Menu

Press Focus to open the Focus Command/Calibrations menu.

Step 3

Turn On Calibration Mode

Press the Calibration button so it reads ON.

Step 4

Select Galvo Platform

Press Galvo Platform to open the Galvo Platform calibration.

Step 5

Start the Calibration

Clear the worktable, then press Start.

Step 6

Finish the Calibration

Let the engraver complete the calibration. When Calibration done appears, press Done.


Fiber Focus Calibration

This calibration sets the focal height of the fiber laser and the manual focus gauge. You will run a small job while adjusting the galvo platform to find the correct focal height.

Step 7

Create the Test Job

In your preferred illustrating program, create a black raster box about 2″ x 2″ (50 mm x 50 mm), and send it to the Software Suite.

Step 8

Set the Job Settings

Set the Process Type to Hatch and use these settings:

  • Speed: 100
  • Power: 40
  • Wave: 10

Step 9

Run the Job

Send the job to the engraver. Place the anodized aluminum in the middle of the worktable, close the door, and press Go to start the job.

Step 10

Press Focus

While the job is running, press the Focus button (circled in red).

Step 11

Find the Focal Height

Use the joystick to slightly raise and lower the galvo platform while watching the spark where the laser hits the aluminum. The correct focal height is where the spark is brightest.

Step 12

Stop the Job

Let go of the joystick. Press Go/Stop to pause the job, then press Reset (circled in red) to stop it.

Step 13

Open Settings Menu

Remove the anodized aluminum from the engraver, then press the gear icon to open the Settings menu.

Step 14

Open Focus Menu

Press Focus to open the Focus Command/Calibrations menu.

Step 15

Select Fiber Focus

Press Fiber Focus to open the Fiber Focus calibration.

Step 16

Enter the Thickness

Make sure the Thickness box matches the material you just marked. For anodized aluminum supplied by Epilog Laser, enter 0.025.

Step 17

Start the Calibration

Press Start. This calibration is very fast because the engraver simply saves the current position of the galvo platform.

Step 18

Finish the Calibration

Let the engraver complete the calibration. A Calibration done message appears when it is finished.


Red Dot Diode Calibration

Step 19

Turn On the Red Dot

Press the Focus button (circled in red) to turn on the focus cross hair and the focus red dot.

Step 20

Check the Red Dot

With the galvo platform at the correct focal height (a focus reading of 0.00), the red dot should be in the center of the cross hair.

Step 21

Adjust if Needed

If the red dot is not centered, adjust it by loosening or tightening the 1/8″ Allen screw on the underside of the red dot assembly.

Step 22

F163 (4″ x 4″) Lens

If calibrating with the F163 lens installed, adjust the Allen screw shown by the arrow.

Step 23

F254 (6″ x 6″) Lens

If calibrating with the F254 lens installed, adjust the Allen screw shown by the arrow.


Manual Focus Gauge Calibration

Step 24

Check the Manual Focus Gauge

With the galvo platform at the correct focal height (a focus reading of 0.00), place the manual focus gauge in the bracket next to the lens. It should be fully seated in the bracket with its tip touching the worktable.

Step 25

Adjust if Needed

If the gauge is not fully seated or does not touch the table, loosen the locking nut, then turn the adjustment nut to set the length. Tighten the locking nut when the length is correct.


Camera Calibration

Step 26

Open Settings Menu

On the display, press the gear icon to open the Settings menu.

Step 27

Open Advanced Settings

Press and hold the word Settings to open the Advanced Settings menu.

Step 28

Select Calibrate Cameras

Press Calibrate Cameras.

Step 29

Place the Calibration Mat

Place the camera calibration mat in the center of the worktable and hold it flat with four of the red-capped thumbscrews.

Step 30

Start the Calibration

Press OK to begin the overhead camera calibration.

Step 31

Let the Calibration Run

A camera feed appears on the display while the calibration runs. The feed will look skewed or slanted; this is normal and does not affect the calibration.

Step 32

Check the Score

When the calibration finishes, a message shows the calibration score. Press OK.

Note: The score varies with lens type, lighting, and how flat the mat is. 1.2 or less is good, 1.2 – 1.7 is acceptable but marginal, and 1.8 or higher may prompt you to run the calibration again. For a poor score, make sure the focus calibrations were completed first, the mat is flat and undamaged, and nothing else is on the worktable.


Galvo Calibration

Step 33

Place the Anodized Aluminum

Place an 8″ x 8″ piece of anodized aluminum in the center of the worktable and hold it flat with four of the red-capped thumbscrews.

Step 34

Start the Calibration

Press OK to begin. The engraver marks the aluminum, then the cameras inspect the mark and adjust so the galvo and camera positioning systems line up.

Step 35

Check the Score

When the calibration finishes, a message shows the calibration score.

Note: 1.8 or less is good, 1.9 – 2.5 is acceptable but marginal, and higher than 2.5 may prompt you to run the calibration again. For a poor score, make sure the focus calibrations were completed first, the aluminum is flat and has not been marked before, and nothing else is on the worktable.