Overhead Camera Calibration
- Install Mat: Place calibration mat in the upper-left corner of the table. Ensure the green circle is in the corner.
- Run: Close doors and press OK.
- Follow the prompt and close the top door of the engraver.
- Allow the process to run until completion. The camera at the laser head will take pictures of the mat to calibrate itself. This process takes several minutes.
- Once the calibration has successfully finished, a dialog will appear. The “Reprojection error” will be listed and is used to determine the quality of the calibration. The lower the value, the better. A value of 0.6 or lower is desired. If the reprojection error is above 2.0, the user will see an error “Calibration unsuccessful. Error too high”.